91980-3Specific airway resistance by Plethysmograph body boxActive
Part Description
LP310432-2 Specific resistance.airway
Specific airway resistance (sRaw) reflects airway resistance (Raw) that has been adjusted for lung volume. It is measured by plethysmography during normal breathing (i.e., tidal breathing), does not require special maneuvers, and is independent of height and sex, so it is commonly used in pediatrics and can be used for children as young as 2. The relationship between sRaw and Raw is: sRaw = Raw * TGV, where TGV is thoracic gas volume. There are different methods used for calculating sRaw from the data gathered during plethysmography, including:
1. sRtot (total specific resistance), which uses the flow measurements at the maximum change in pressure
2. sRaw Vmax, which uses the the measurements at the maximum flow
3. sRaw0.5, which uses the measurements where the flow reaches 0.5 L/s
Specifying the method used to calculate sRaw is important for interpreting the results.
[PMID:16002957]
Source: Regenstrief LOINC
Fully-Specified Name
- Component
- Specific resistance.airway
- Property
- FldResist
- Time
- Stdy^mean
- System
- Respiratory system.airway
- Scale
- Qn
- Method
- Plethysmograph.body box
Additional Names
- Short Name
- sRaw Plethysmography
Basic Attributes
- Class
- PULM
- Type
- Clinical
- First Released
- Version 2.66
- Last Updated
- Version 2.67
- Change Reason
- Release 2.67: FORMULA: Added alternate formulas based on functional residual capacity measured by plethysmograph body box
- Order vs. Observation
- Both
Formula (Readable)
Multiple calculation methods: sRaw = Raw * TGV; sRaw = Raw * FRCpleth; sRaw = Raw * (FRCpleth + VT/2); where sRaw is specific airway resistance, Raw is airway resistance, TGV is thoracic gas volume, FRCpleth is functional residual capacity measured by plethysmograph body box, and VT is tidal volume.
Language Variants Get Info
- zh-CNChinese (China)
- 特异性阻力.气道:
液体阻力: 检查过程持续时间^平均值: 呼吸系统.气道: 定量型: 体积描记器.体箱计 - it-ITItalian (Italy)
- Resistenza specifica.vie aeree:
FldResist: Studio^media: Apparato respiratorio.vie aeree: Qn: Pletismografo.body box - es-MXSpanish (Mexico)
- Resistencia específica vía aérea:
FldResistenciat: Estudio^prom: Sistema respiratorio vía aérea.: Cuantitativo: Plethysmograph.body box
Example Units
Unit | Source |
---|---|
cm[H2O].s | Example UCUM Units |
LOINC FHIR® API Example - CodeSystem Request Get Info
https://fhir.loinc.org/CodeSystem/$lookup?system=http://loinc.org&code=91980-3
LOINC Copyright
Copyright © 2022 Regenstrief Institute, Inc. All Rights Reserved. To the extent included herein, the LOINC table and LOINC codes are copyright